⚠️ Common HPLC Problems & Troubleshooting
1. High Back Pressure 🚧
Possible causes
Blocked column
Dirty mobile phase
Blocked inline filter
Precipitated buffer salts
Troubleshooting
Flush column with strong solvent (e.g., high % organic)
Replace guard column or inline filter
Filter and degas mobile phase (0.45 µm or 0.22 µm)
Check tubing for blockage
---
2. Low Pressure 📉
Possible causes
Leak in system
Pump seal worn out
Empty solvent reservoir
Air in pump
Troubleshooting
Check for leaks in fittings
Prime the pump properly
Replace pump seals
Ensure solvent bottle has enough mobile phase
---
3. Baseline Noise 🔊
Possible causes
Air bubbles in detector
Dirty flow cell
Poor degassing
Contaminated mobile phase
Troubleshooting
Degas mobile phase (sonication or helium sparging)
Clean detector flow cell
Purge detector
Use fresh solvents
---
4. Baseline Drift 📈
Possible causes
Temperature fluctuations
Mobile phase composition change
Column not equilibrated
Detector lamp aging
Troubleshooting
Allow system to equilibrate (20–30 min)
Maintain constant column temperature
Prepare fresh mobile phase
Check detector lamp intensity
---
5. Peak Tailing 🧪➡️
Possible causes
Column contamination
Secondary interactions
Incorrect pH
Dead volume in fittings
Troubleshooting
Flush column with strong solvent
Adjust mobile phase pH
Replace column if old
Tighten fittings properly
---
6. Peak Fronting ⬅️🧪
Possible causes
Column overload
Injection volume too high
Sample solvent stronger than mobile phase
Troubleshooting
Reduce injection volume
Dilute sample
Match sample solvent with mobile phase
---
7. No Peaks Detected ❌
Possible causes
Wrong wavelength
No injection
Detector off
Sample degradation
Troubleshooting
Check detector wavelength
Confirm injector working
Check sample preparation
Verify standard solution
---
8. Ghost Peaks 👻
Possible causes
Contaminated mobile phase
Carryover from previous injection
Dirty injector
Troubleshooting
Use fresh solvents
Wash injector thoroughly
Run blank injection
Clean autosampler needle
---
9. Retention Time Shift ⏱️
Possible causes
Flow rate variation
Column temperature change
Mobile phase composition change
Troubleshooting
Calibrate pump flow rate
Maintain column oven temperature
Prepare mobile phase accurately
---
10. Irregular Peak Shape 📉📈
Possible causes
Air bubbles
Column void
Damaged column packing
Troubleshooting
Degas mobile phase
Reverse flush column (if allowed)
Replace column
---
🧰 Quick Daily Preventive Tips
Filter mobile phase 🧴
Degas solvents 💨
Flush system after buffer use 🔄
Use guard column 🛡️
Maintain logbook 📒