⚠️ Common HPLC Problems & Troubleshooting 1. High Back Pressure ๐ง Possible causes Blocked column Dirty mobile phase Blocked inline filter Precipitated buffer salts Troubleshooting Flush column with strong solvent (e.g., high % organic) Replace guard column or inline filter Filter and degas mobile phase (0.45 ยตm or 0.22 ยตm) Check tubing for blockage --- 2. Low Pressure ๐ Possible causes Leak in system Pump seal worn out Empty solvent reservoir Air in pump Troubleshooting Check for leaks in fittings Prime the pump properly Replace pump seals Ensure solvent bottle has enough mobile phase --- 3. Baseline Noise ๐ Possible causes Air bubbles in detector Dirty flow cell Poor degassing Contaminated mobile phase Troubleshooting Degas mobile phase (sonication or helium sparging) Clean detector flow cell Purge detector Use fresh solvents --- 4. Baseline Drift ๐ Possible causes Temperature fluctuations Mobile phase composition change Column not equilibrated Detector lamp aging Troubleshoot...
The Pharmaceutical Study and Latest Update Blog is a platform for sharing insights into the latest studies and developments in the field of pharmaceutical research. The blog aims to keep readers informed about the latest advancements in drug discovery, clinical trials, and regulatory approvals. It also provides a platform for discussion and debate on the latest trends in pharmaceutical science. In addition to providing updates on the latest breakthroughs in drug development, the blog also cover